All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.

Features for Micropette:

* Light weight, ergonomic, low force design

* Digital display clearly reads volume setting

* The pipettes cover volume range of 0.1μl to 10ml

* Easy to calibrate and maintain

* The lower part is autoclavable

* Design helps avoid repetitive strain injuries

* Manufactured from innovative materials

Features for Multi-channel Micropette:

* 8 and 12 channel pipettes are available for standard 96-well plate

* Dispensing head rotates ensure optimum pipetting convenience

* Individual piston/tip cone assemblies allow easy repair and maintenance

* Compound material tip cone design allows visual seal vertification

* Universal pipette tips can be used

8-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume

(μl)

Maximume Permissible Systematic error

(Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

% μl % μl
0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4