Fluke Calibration

Fluke Calibration 3130: Portable Pressure Calibrator

Download         Download

             00961 1 894 243

Description

Features

  • Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar)
  • Internal pump can generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar)
  • Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar)
  • Includes variable volume for fine adjustment of pressures
  • Pressure measurement accuracy of ± (0.025 % reading + 0.01 % FS)
  • Electrical measurement and 24 volt supply for close looped calibrations
  • Measure or generate 4 to 20 mA
  • Measure 0 to 30 V dc
  • Powered by internal, high capacity, NiMH battery or universal ac mains adapter
  • Compatible with Fluke 700P and 750P Pressure Modules

Specifications:

Environmental
Operating temperature -10°C to +50°C
Storage Temperature -20°C to +60°C
Power requirements 12 V DC (Universal ac adapter/charger supplied)
Battery Internal 3800 mAh advanced NiMH pack
Operating Time
On full charge Approx 50 hours (Measure only or external air; no pump)
Using internal pump Approx 100 calibration cycles to 300 psi
Physical
Dimensions (L x W x D) 15.25 x 12 x 7 in
Weight ~7 kg (15 lb)
EMI/RFI conformance EN61326:2006 Annex A
Connectors/ports 1/8 in NPT (External supply port and test port)
Included accessories Manual, NIST-traceable certificate, test leads, universal ac adapter/charger
Ranges
Pressure (internal pump) -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar)
Pressure (external air) 0 to 2 MPa (0 to 300 psi, 0 to 20 bar)
mA 0 to 24.000 mA
Volts 0 to 30.000 V DC
Engineering units psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4°C, mmH2O @ 20°C, cmH2O @ 4°C, cmH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C
Instrumental Measurement Uncertainty
Pressure 0.025% of reading ±0.01% FS
mA 0.015% of reading ±0.002 mA
Volts 0.015% of reading ±0.002 V
Temperature Effect (All Functions)
No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C

Additional information

Labels