FLUKE

FLuke 5624-20-X: Probe, 1000°C, 10 ohm PRT, 6.35 mm x 508 mm (0.25″ x 20″), (includes 2609 case)

Download

             00961 1 894 243

Category:

Description

  • Fast Response PRT, 3.2 mm (0.13 in), lead wire (cable) 6 ft
  • 5624 Platinum Resistance Thermometer with Temperature Range of 0 °C to 1000 °C
    • Temperature range of 0°C to 1000°C
    • Accuracy (includes short-term stability and calibration uncertainty) of ±0.05°C to 962°C
    • Long-term drift of 0.01°C at 0°C after 100 hours at 1000°C
    • Designed by Fluke Calibration’s primary standards design team

      Specifications

      Range 0 °C to 1000 °C
      Calibration Uncertainty
      ± 0.004 °C at 0 °C
      ± 0.020 °C at 962 °C
      Long-term Drift (Rtpw) (k=2)
      <0.01 °C at 0 °C/100 hours at 1000 °C
      <0.06 °C at 0 °C/1000 hours at 1000 °C
      Calibrated Accuracy† (k=2)

      ± 0.006 °C at 0 °C
      ± 0.031 °C at 660 °C
      ± 0.041 °C at 961 °C
      Short-term Stability ± 0.005 °C
      Immersion <153 mm (6 in) at 700 °C
      Rtpw 10 ohm (± 0.1 ohm)
      Hysteresis <0.005 °C from 0 °C to 1000 °C
      Thermocycling <0.01 °C, 10 cycles from 0 to 1000 °C
      Current 1 mA
      Size 6.35 mm (0.25 in) O.D.
      Length 508 mm (20 in)
      Sheath Material Alumina
      Calibration Included 1913-6 fixed-point calibration

Additional information

Labels