Description
Analyze airflow. Easily. The 922 Micromanometer.
Today’s HVAC technicians want a simple solution for diagnosing ventilation issues. Differential pressure measurements only tell part of the story. Technicians also want to measure air velocity and flow, without having to resort to expensive, difficult to use, specialist tools. The Fluke 922 makes airflow measurements easy by combining three tools: differential pressure, airflow, and velocity into a single, rugged meter.
Use the Fluke 922 micromanometer to:
|
Technical Specifications | ||
Air pressure | Range / Resolution / Accuracy | ±4000 Pascals / 1 Pascal / ±1% + 1 Pascal |
±16 inH2O / 0.001 inH2O / ±1% + 0.01 inH2O | ||
±400 mmH2O / 0.1 mmH2O / ±1% + 0.1 mmH2O | ||
±40 mbar / 0.01 mbar / ±1% + 0.01 mbar | ||
±0.6 psi / 0.0001 psi / ±1% + 0.0001 psi | ||
Air velocity | Range / Resolution / Accuracy | 250 to 16,000 fpm / 1 fpm / ±2.5% of reading at 2000 fpm (10.00 m/s) |
1 to 80 m/s / 0.001 m/s / ±2.5% of reading at 2000 fpm (10.00 m/s) | ||
Air flow (volume) | Range / Resolution / Accuracy | 0 to 99,999 cfm / 1 cfm / Accuracy is a function of velocity and duct size |
0 to 99,999 / 1 m3/hr / Accuracy is a function of velocity and duct size | ||
0 to 99,999 l/s / 1 l/s / Accuracy is a function of velocity and duct size | ||
Temperature | Range / Resolution / Accuracy | 0°C to 50°C / ±1% + 2°C / 0.1°C |
32°F to 122°F / ±1% + 4°F / 0.1°F | ||
General Specifications | ||
Operating temperature | 0°C to +50°C (+32°F to +122°F) | |
Storage temperature | -40°C to +60°C (-40°F to +140°F) | |
Relative humidity | 0% to 90%, non-condensing | |
IP rating | IP40 | |
Operating altitude | 2000 m | |
Storage altitude | 12000 m | |
EMI, RFI, EMC | Meets requirements for EN61326-1 | |
Vibration | MIL-PREF-28800F, Class 3 | |
Max pressure at each port | 10 psi | |
Data storage capacity | 99 readings | |
Warranty | 2 years | |
Power, battery life | Four AA batteries | |
Battery life | 280 hours without backlight, 60 hours with backlight |