Fluke Biomedical ProSim SPOT Light: SpO2 Pulse Oximeter Analyzer

Download         Download

             00961 1 894 243

Category:

Description

ProSim SPOT Light
Featuring an exclusive ergonomic design, the ProSim SPOT Light is the first comprehensive SpO2 functional tester to come in a handheld easy-to-use device. SPOT Light is lightweight and flexible, with three custom presets specially-designed to make it the fastest and easiest-to-use device on the market today for pulse oximeter functional testing. A helpful LCD display and three simple buttons make it effortless to rapidly change parameters and view each signal output sent to the pulse oximeter at a glance. An interchangeable, long-life battery ensures uninterrupted all-day operation without the need to connect to a power supply.
SPOT Light sets up in seconds to send SpO2 saturation, heart rate, perfusion, transmission, artifact noise, and eight different manufacturers custom R-curves to a pulse oximeter or patient monitor. It is designed to take the abuse of everyday use, including the stress of transportation, moving from room to room and device to device, and the occasional fall from the workbench.
An increasing number of medical devices now have pulse oximetry built in. SPOT Light can be paired with other Fluke Biomedical test tools to make modern device testing task fast and easy. Need to check a defibrillator with pulse oximetry? SPOT Light with the Impulse 7000 Defibrillator Analyzer allow you to test any external defibrillator on the market today. Do you have a fleet of new patient monitors coming that require incoming inspection? Using SPOT Light with ProSim 4 Vital Signs Simulator enables you to easily accomplish this task in hours instead of days.
Key features:
Small, portable, and lightweight, weighing in at 8.8 oz
Large, easy to read LCD display
Rechargeable battery lasts 10 hours minimum
Signal strength indicator
SpO2 Saturation 80 % to 100 %
Heart rate: 30 BPM to 240 BPM
Perfusion: 0.2 %, 2 % and 10 %
Transmission: dark/thick, normal and light/thin
Artifacts: respiration and ambient light
Three editable test sequences
R-curves: 8 including Masimo, Nellcor and Nonin
Expandable pairing capabilities

Additional information

Labels