Description
| Ordering Code | PLUS2 |
| Power Mode | |
| Power Range | 100 nW – 10 kW |
| Resolution | 0.01% – 0.1% of Full Scale |
| Energy Mode | |
| Energy Range | 1 mJ – 1 kJ |
| Resolution | 0.01% – 0.1% of Full Scale |
| Pulse repetition rate Range | Single shot – 6 Hz (a) |
| Power Probe Mode (FIT-H) | |
| Power Range | 100 mW – 10 kW |
| Resolution | 0.5 ‰ (a) |
| General Operating Characteristics | |
| Sensor Compatibility | Thermal sensors, OEM thermal sensors, FIT-H power probes, Photodiodes, Blink FR |
| Storage Temperature Range | -20 / +70 °C |
| Operating Temperature Range | +5 / +40 °C |
| Relative Humidity Range | 10 – 70 % |
| PC connectivity | USB |
| Sensor Head connectivity | DB15 |
| Selectable Wavelengths | 1 nm resolution (b) |
| Dimensions | 170 H x 100 W x 36-50 D (mm) |
| Weight | 380 g |
| Display | 4.3” TFT LCD high brightness; 480 x 272 resolution; resistive touch screen; 96 H x 55 W mm |
| Battery | Built in rechargeable Li-Pol; 3.7 V, 3700 mAh |
| Battery charge time | 7-8 hours if not working; 15-20 hours if working |
| Battery run time | >9 hours in normal operations; >15 hours in stand-by display mode |
| Supplied battery charger | Input 100/240 Vac; 50/100 Hz; 5 Vdc; 1A; Charging current: 0.5 A; (Plus2 may be charged through a PC USB port) |
| Electronics Characteristics | |
| ADC sampling rate | 64 Hz – 192 Hz (c) |
| ADC resolution | 23 bit |
| Electrical accuracy | ± 0.5 % |
| Analog Output | 0.025 – 2 V (with 16 bit resolution; 0.0015% resolution) |
| Analog Output Accuracy | ± 0.1%; ± 2 mV relative to display |
| Notes | |
| a) for Blink FR b) available wavelength range is sensor head dependent c) sampling frequency is sensor head dependent |
|